Transgenomic ICSep AN1
Transgenomic ICSep AN1 Part Number: ANX-99-8511
Description
Size: 4.6 x 250mm PEEK
Bead Size: 7um
Notes:
- Strong Anion Exchange column, Polystyrene Divinylbenzene
- Can be used with Non-Suppressed Conductivity Detection
- Very good Fluoride separateion from Water Dip
- Great for Tooth paste analysis
- Great for Trace anion work
- Great for Speciation (Chromium and Selenium, etc)
Chromatogram:

Analysis Conditions:
- Eluent: 1.7mM Sodium Carbonate / 1.8mM Sodium Bicarbonate
- Flow Rate: 1.0mL/min
- Temperature: Ambient
- Detection: Suppressed Conductivity
- Injection: 20uL
Sample:
| Peak # | Analyte | Retention Time | Sample Concentration |
| 1 | Fluoride | 4.00 min | 5 ppm |
| 2 | Chloride | 5.06 min | 10 ppm |
| 3 | Nitrite | 5.92 min | 10 ppm |
| 4 | Bromide | 7.25 min | 10 ppm |
| 5 | Nitrate | 8.35 min | 20 ppm |
| 6 | Phosphate | 11.94 min | 30 ppm |
| 7 | Sulfate | 14.27 min | 20 ppm |